Overview
Absorbance quantifies the attenuation of light passing through a sample at a selected wavelength. UV–visible spectroscopy measures this response in the ultraviolet and visible regions. In life science laboratories, absorbance measurements support nucleic acid quantification, protein assays, enzyme kinetics and colorimetric microplate assays. [1]
The Beer Lambert relationship
Absorbance is defined as A = log10(I0/I), where I0 and I represent reference and transmitted light intensities. For an appropriate homogeneous absorbing solution, the Beer–Lambert relationship is A = εlc: molar absorptivity multiplied by optical path length and concentration. Absorbance is dimensionless. The relationship applies within suitable chemical and instrumental conditions; concentration should be calculated only over a validated measurement range. [1,2]
Why path length matters in microplates
In a microplate well, the optical path depends on the liquid height and well geometry. Changing sample volume can therefore change absorbance even when concentration remains constant. Use controlled volumes, a validated standard curve or an appropriate path-length correction method. Microvolume accessories, like the Take 3 microvolume plates, use defined short optical paths to measure small samples; their calculations must account for the actual measurement geometry. [2]
Direct measurements and colorimetric assays
Nucleic acids can be measured through their UV absorbance, while assays such as BCA, Bradford and ELISA generate or measure coloured species. Each method has its own wavelength, blanks, standards and interference profile. UV measurements require suitable UV-transparent optical surfaces. Microbial optical density is a related transmission measurement, but light scattering by cells contributes strongly. Its relationship to cell concentration needs validation and should not be treated as simple molecular absorbance.
Sources of measurement error
Blank the measurement with a suitable matrix and reagents. Bubbles, particles, fingerprints, precipitation and plate background can affect light transmission. Strongly absorbing samples can exceed the useful linear range and may require dilution. Buffer components or contaminants can also absorb at the analytical wavelength. Absorbance ratios used for nucleic acid quality assessment are indicative and cannot uniquely identify contaminants or establish suitability for every downstream assay.
Selecting an instrument

Common questions
Does a higher absorbance always mean a higher analyte concentration?
Only when path length, chemistry and background are comparable and the measurement remains within its validated range. Scattering and interfering compounds can also raise the signal.
Can every plastic microplate be used at 260 nm?
No. Use labware with optical properties appropriate for the selected UV wavelength and confirm compatibility with the reader.
Need help selecting a platform for your assay? Contact BioSPX to discuss the sample, readout and configuration you need.
References
[1] Agilent Absorbance technology