Hybrid plate reader optics

Your experts in plate readers in Belgium, Luxembourg and the Netherlands

Hybrid plate reader optics explained

Hybrid plate reader optics combine filter-based and monochromator-based wavelength selection within one instrument. Agilent BioTek uses the name Hybrid Technology for this arrangement in compatible multimode readers. The two optical approaches support different assay requirements, allowing laboratories to select the appropriate path for a measurement. [1]

How filter based optics work

Filter-based optics select the wavelength bands used to excite a fluorophore and detect its fluorescence. In Agilent BioTek hybrid readers, excitation and emission bandpass filters work together with dichroic mirrors to direct light through the optical system. High transmission within the selected bands, combined with strong blocking outside them, helps deliver excitation light to the sample while limiting unwanted light reaching the detector.

Filters can be selected to match the spectral properties of specific fluorophores, making this approach well suited to established assays with defined wavelength requirements. Measuring a different fluorophore may require another filter set or filter cube if its excitation or emission falls outside the installed bands.

Assay performance depends on the complete optical system and measurement conditions, including filter characteristics, detector settings and sample background. [1,2]

How monochromator based optics work

A monochromator separates light into its component wavelengths and selects a defined spectral band for excitation or detection. Agilent BioTek hybrid readers use a quadruple-grating monochromator design, with wavelength selection controlled through the instrument software. Researchers can adjust excitation and emission settings within the supported range without installing a dedicated optical filter for each wavelength.

This flexibility is useful when optimising detection conditions for different fluorophores or developing new assays. Excitation and emission scans can help locate fluorescence peaks and select suitable measurement wavelengths under the assay conditions.

Configurations with variable bandwidth also allow adjustment of the width of the selected spectral band. Wider bands can increase the collected signal, but may also admit more background or overlapping fluorescence. Narrower bands improve spectral selectivity while transmitting less light. The most suitable settings therefore depend on the fluorophore, sample background and required assay performance. [1,3]

Why combine both approaches

A laboratory developing new assays may value flexible wavelength selection. An established assay may benefit from a dedicated filter path. Having both paths available can support those workflows on one platform, provided the necessary modules are installed. Hybrid describes the optical design; multimode describes the availability of different measurement modes. A multimode reader is therefore not necessarily a hybrid reader.

Evaluating assay performance

Compare signal-to-background ratio, variability, useful dynamic range and detection limits using the actual assay. Avoid assuming that filter optics are always more sensitive or that a monochromator will always provide better specificity. Wavelengths, bandwidth, excitation source, detector, read geometry and integration time all contribute. Assay validation should use the settings intended for routine experiments rather than a generic instrument specification.

Configuration and upgrade considerations

BioSPX lists Synergy H1 and Synergy Neo2, both offered with hybrid configurations. Available modes and optical modules vary by model and configuration. Confirm whether the required assay needs fluorescence intensity, fluorescence polarization, time-resolved detection or another mode, and whether the installed hardware supports it. Additional modules or accessories may be required; the term hybrid alone does not establish support for every assay. [2,3]

Relevant BioSPX instruments

Do you need help selecting a platform for your assay? Contact BioSPX to discuss the sample, readout and configuration you need.

FAQ

The term indicates that both approaches are available in the instrument. How they are used depends on the measurement mode and protocol.

No. Step size describes the increment between selected wavelengths. Bandwidth describes the range of wavelengths passed at a selected setting.